JUMO's calibration services: where precision and reliability come together. Accuracy is crucial in your world of measurements. That is why we provide high-quality calibration with a certificate, so that you can always rely on the quality of your instruments. On this page, we show you how our services support you. We highlight the accreditation bodies that set the standards worldwide: BELAC (Belgium), DAkkS (Germany) and the Dutch Accreditation Council (RvA) in the Netherlands. This way, you can be sure that your calibration is internationally recognised, giving you peace of mind!
Calibration of relative pressure according to DKD R 6-1 (German calibration service) | |
---|---|
Measuring range -1 to -0.4 bar | Extended measurement uncertainty 2 mbar (1) |
Measuring range > -0.4 to 0.4 bar | Extended measurement uncertainty 0.4 mbar (1) |
Measuring range > 0.4 to 2 bar | Extended measurement uncertainty 2 mbar (1) |
Measuring range > 2 to 6 bar | Extended measurement uncertainty 4 mbar (1) |
Measuring range > 6 to 20 bar | Extended measurement uncertainty 10 mbar (1) |
Measuring range > 20 to 60 bar | Extended measurement uncertainty 30 mbar (1) |
Measuring range 0 to 600 bar | Extended measurement uncertainty 0.35 mbar (2) |
Pressure medium: (1) gas, (2) water |
Calibration absolute pressure according to DKD-R 6-1 (German calibration service) | |
---|---|
Measuring range 0.7 to 1.2 bar | Extended measurement uncertainty 0.5 mbar (1) |
Measuring range > 0 to 3 bar | Extended measurement uncertainty 2 mbar (1) |
Measuring range > 3 to 7 bar | Extended measurement uncertainty 4 mbar (1) |
Measuring range > 7 to 21 bar | Extended measurement uncertainty 10 mbar (1) |
Measuring range > 21 to 61 bar | Extended measurement uncertainty 30 mbar (1) |
Measuring range 1 to 601 bar | Extended measurement uncertainty 0.35 mbar (2) |
Pressure medium: (1) gas, (2) water |
As an accredited laboratory, our measurement results can be traced back to national and international standards, so that the issued calibration certificate creates international comparability and confidence for the end user. At the same time, the calibrations performed have only minimal measurement uncertainties, guarantee the reliable measurement of pressure measuring devices, and make a significant contribution to process reliability.